Woven EndoBridge (WEB)

Woven EndoBridge (WEB)

The Woven EndoBridge (WEB) (Sequent Medical, Aliso Viejo, California), is a ellipsoid braided-wire embolization device designed to provide flow disruption along the aneurysm neck 1).

Placed in the aneurysm, the device will modify the blood flow at the level of the neck and induce aneurysmal thrombosis. The WEB shape was designed to treat wide necked aneurysm. The device has been developed progressively from a dual-layer version (WEB DL) to single-layer versions (WEB SL and WEB SLS [single-layer spherical]).

This device does not require long-term antiplatelet use.

For the treatment of both ruptured and unruptured aneurysms. The WEB has received the CE mark and to date has been used to treat a wide variety of more than 1,400 aneurysms in Europe, Latin America and New Zealand. The WEB is not available for sale or use in the United States.

The WEB is a self-expanding, oblate, braided nitinol mesh.

The device is composed of an inner and outer braid held together by proximal, middle, and distal radiopaque markers, creating 2 compartments: 1 distal and 1 proximal. Depending on the device diame- ter, the inner and outer braids are 108 wires or 144 wires. Therefore, blood flow into a WEB-embolized aneurysm initially encounters 2 layers of wires comprising 216 or 288 wires, with the largest interwire distance ranging from 106 to 181 􏰅m, respectively, depending on the device size. The WEB implant is deployed—or retrieved before de- tachment—in a manner similar to that in endovascular coil systems, through microcatheters with an internal diameter 􏰆0.027 inch. For devices with a diameter of 􏰇7 mm, microcatheters with an internal diameter of 0.027 inch are used; and for devices with a diameter 􏰁7 mm, microcatheters with an internal diameter 0.032 inch are used. The detachment system is electrothermal and instantaneous. 2).


In a study, there was no difference in the early clinical course between those treated with WEB embolization, coil embolization, or neurosurgical clipping. Since WEB embolization is a valuable treatment alternative to coiling, it seems not justified to exclude this procedure from upcoming clinical SAH trials, yet the clinical long-term outcome, aneurysm occlusion, and retreatment rates have to be analyzed in further studies 3).

WEB Intrasaccular Therapy Study (WEB-IT)


The WEB Clinical Assessment of Intrasaccular Aneurysm Therapy (WEBCAST) trial is a prospective European trial evaluating the safety and efficacy of WEB in wide necked aneurysm of the bifurcation.

>

It does not immediately secure the aneurysm in most subarachnoid hemorrhage cases. Second, it may not be suitable for embolization of wide-neck aneurysms with an unfavorable aspect ratio. To overcome these limitations, Zanaty et al., used the WEB device in conjunction with stenting and/or coiling.

They presented a technical note with an illustrated case-series, and provide a detailed step-by-step description on how the WEB device can be used in adjunct to coiling and/or stenting to achieve successful angiographic results. Accurate sizing of the WEB device before deployment is critical. Larger case-series are required to further assess the safety and success of these combined techniques 4).

Zhang et al. searched the PubMedOvid MEDLINE, and EMBASE databases between December 1, 2012 and June 30, 2018.

Studies that included five or more patients undergoing WEB for Wide necked intracranial aneurysms, reported an angiographic or clinical outcome and risk factors, and were published after December 1, 2012 were eligible.

Major outcomes included initial or short-term complete and adequate occlusion. Secondary outcomes included treatment failure, recanalizationmortalitymorbidity, and complication (e.g., thromboembolism or intraoperative rupture) rates. A random-effect model was used to pool the data. To assess risk factors for short-term angiographic outcomes and the most common complications, they conducted subgroup analyses and obtained odds ratios with 95% confidence intervals.

They included 36 studies (1759 patients with 1749 aneurysms). The initial complete and adequate occlusion rates were 35% and 77%, respectively. After a mean follow-up of 9.34 months, the short-term complete and adequate occlusion rates were 53% and 80%, respectively. Thromboembolism and recanalization were the most common complications (both 9%), followed by mortality (7%), morbidity (6%), failure (5%) and intraoperative rupture (3%). The following factors were related to higher short-term obliteration rates: unruptured status, in the anterior circulation, a medium neck (4-9.9 mm), newer-generation WEB and treatment without additional devices. Ruptured status, anterior circulation, preoperative antiplatelet therapy and newer-generation WEB were not significantly related to withto thromboembolism.

WEB has a satisfactory safety profile and shows promising efficacy in treating wide-neck intracranial aneurysms. They preliminarily identified several risk factors for short-term angiographic outcomes 5).

Woven EndoBridge case series.

A case of a shallow basilar tip aneurysm treated with the WEB device that required stabilization with Y-stent through radial access 6).


1)

Ding YH, Lewis DA, Kadirvel R, Dai D, Kallmes DF. The Woven EndoBridge: a new aneurysm occlusion device. AJNR Am J Neuroradiol. 2011 Mar;32(3):607-11. doi: 10.3174/ajnr.A2399. Epub 2011 Feb 17. PubMed PMID: 21330397.
2)

Pierot L, Liebig T, Sychra V, Kadziolka K, Dorn F, Strasilla C, Kabbasch C, Klisch J. Intrasaccular flow-disruption treatment of intracranial aneurysms: preliminary results of a multicenter clinical study. AJNR Am J Neuroradiol. 2012 Aug;33(7):1232-8. doi: 10.3174/ajnr.A3191. Epub 2012 Jun 7. PubMed PMID: 22678844.
3)

Sauvigny T, Nawka MT, Schweingruber N, Mader MM, Regelsberger J, Schmidt NO, Westphal M, Czorlich P. Early clinical course after aneurysmal subarachnoid hemorrhage: comparison of patients treated with Woven EndoBridge, microsurgical clipping, or endovascular coiling. Acta Neurochir (Wien). 2019 Jul 6. doi: 10.1007/s00701-019-03992-4. [Epub ahead of print] PubMed PMID: 31280480.
4)

Zanaty M, Roa JA, Dandapat S, Samaniego EA, Jabbour P, Hasan D. Diverse Use of the WEB Device: A Technical Note on WEB Stenting and WEB Coiling of Complex Aneurysms. World Neurosurg. 2019 Jul 10. pii: S1878-8750(19)31933-3. doi: 10.1016/j.wneu.2019.07.027. [Epub ahead of print] PubMed PMID: 31301439.
5)

Zhang SM, Liu LX, Ren PW, Xie XD, Miao J. Effectiveness, safety and risk factors of Woven EndoBridge device in the treatment of wide-neck intracranial aneurysms : systematic review and meta-analysis. World Neurosurg. 2019 Aug 13. pii: S1878-8750(19)32175-8. doi: 10.1016/j.wneu.2019.08.023. [Epub ahead of print] PubMed PMID: 31419591.
6)

Nordmann NJ, Weber MW, Dayoub H. Woven Endobridge (WEB) augmented by Y-stent in a shallow basilar tip aneurysm. J Cerebrovasc Endovasc Neurosurg. 2022 Feb 17. doi: 10.7461/jcen.2022.E2021.08.006. Epub ahead of print. PMID: 35172073.

Leave a Reply